Laboratory tests
Real data from the laboratory model
Before flying, the design is tested in the laboratory. What follows are results from the development model of the digital processor and one of the two analog boards, running the final test version of its software.
Continuous run in the longest observation mode. A single analog board, with the antenna input connected to the test bench, not to a flight antenna.
A real spectrum
Spectrum measured by the instrument on the bench, from 0 to 250 MHz: the internal reference load (the shape of the receiver’s response) and the bench antenna input, which picks up signals from the laboratory environment. The science band starts at 50 MHz. The shaded band is the range between the 10th and 90th percentiles of the values in each channel.
Arbitrary processor units, logarithmic scale. Median of each channel over the whole observation. The narrow peaks on the antenna input are signals from the bench environment.
More than 170 hours non-stop
Temperature of the processor and the analog board during the run. After the initial warm-up, the processor stayed stable between about 57 and 61 °C, with no resets.
One telemetry query every 2 minutes. The marks at the bottom show the arrival of each complete data sequence.
Stability of the levels
How much the signal level of each step changed relative to the first sequence. The internal reference load stayed within 0.5 %. The bench antenna input varies more, because it depends on what it receives.
What was verified
VERIFIEDGain stability: 0.07 dB in 50 hours
The requirement is 0.25 dB. The internal reference load was used throughout the test.
VERIFIEDNo data lost
Every sample goes into a spectrum. In 2-hour steps, 14,400 integrations were counted, as many as the planned duration, and none was lost.
VERIFIEDRecovery from an induced fault
With the digital processor’s clock out of range for 15 seconds, the instrument detected it, restarted the spectrometer and carried on: it kept 191 of 240 integrations and the other steps were not affected.
VERIFIEDTelemetry during observation
The instrument status was requested every 30 seconds while it observed. It answered 257 of 257 requests, in 263 to 298 milliseconds, without altering the data.
VERIFIEDIntegrity of every frame
All frames of the long run arrived with the correct checksum; the last one is partial because the run was stopped. Automated tests also compare the frame byte by byte with an independent model.
VERIFIEDInstrument software tested without hardware
142 deliberate errors were introduced into the frame and sequence code, and the tests caught every one.
VERIFIEDStandalone start-up from non-volatile memory
At power-on the board starts on its own from its non-volatile memory and is ready to observe in about 2 seconds. The memory holds three copies of the program: with two of them deliberately damaged, it started from the third.
VERIFIED20 hours in a row from non-volatile memory
Started from that memory, the board observed for 20 hours and 44 minutes without restarting: all 40 frames arrived with the correct checksum and no data was lost.
What is still to be tested. Behavior over temperature and supply voltage; environmental tests for vibration, thermal vacuum, radiation and electromagnetic compatibility; operation with both analog boards at once; and the antenna in its final version.